Table 2—CS 9600 Technical Specifications
|
Components |
X-ray Generator |
|
Tube voltage |
60 – 90 kV 60 – 120 kV (optional) |
|
Tube current |
2 - 15 mA |
|
Frequency |
140 kHz |
|
Tube focal spot (IEC 60336) |
0.3 or 0.7 mm |
|
Total filtration |
> 2.5 mm eq. Al |
|
Panoramic Modality |
|
|
Sensor technology |
CMOS |
|
Image field |
6.4 x 140 mm (for adult patient size) 6.4 x 120 mm (for child patient size) 120 x 140 mm (for sinus one-shot exam) |
|
Gray scale |
16384 - 14 bits |
|
Magnification |
1.28 |
|
Radiological exams |
Full panoramic Segmented panoramic Bitewing Maxillary sinus Lateral TMJ x 2 Lateral TMJ x 4 Sinus AP / PA / Lateral Orthogonal panoramic |
|
Exposure time |
0.5 to 13 s |
|
Exposure mode |
4 patient sizes (child, small adult, medium adult, large adult) 3 dental arch morphology (normal, square, sharp) |
Note: Panoramic modality is not available in Australia when the device is installed only with the 3D modality.
|
3D Modality |
|
|
Technology |
Dental Volumetric Reconstruction (DVR) |
|
Sensor technology |
CMOS |
|
Volume Field Of View (FOV) diameter x height (cm) |
4 x 4 5 x 5 5 x 8 6 x 6 8 x 5 8 x 8 10 x 51 10 x 10*1 12 x 51 12 x 10*1 16 x 61 16 x 10*1 16 x 121 16 x 17*1 *with tip of the volume |
|
Radiological exams |
Tooth / Teeth Full, upper or lower jaw TMJ Face ENT Upper cervical spine Wrist |
|
Gray scale |
16384 - 14 bits |
|
Magnification |
1.4 |
|
Voxel Size |
75 µm minimum |
|
Exposure time |
5.5 to 40 s (2 x 20 s) |
|
Scan mode |
Continuous |
1In Ontario (Canada), the use by dentists of FOVs that are over 8x8 is subject to conditions.
|
Cephalometric modality (available with Scan Ceph configuration) |
|
|
Sensor technology |
CMOS |
|
Image field |
6.4 x 263.3 mm |
|
Grey scale |
16384 – 14 bits |
|
Magnification |
1.13 |
|
Radiological exams |
Lateral Frontal AP or PA Oblique Submento-vertex Carpus (optional) |
|
Exposure time |
2.96 to 10s |
Note: Cephalometric modality is not available in Australia when the device is installed only with the 3D modality.
|
Components |
CS 9600 |
|
Input voltage (AC) |
100-240 V – 50/60 Hz |
|
Unit dimensions |
1284 (L) x 1669 (D) x 2526 mm (H) |
|
Required space |
1500 (L) x 2000 (D) x 2200 (H) mm (without optional Stool or when Stool installed on the left) 1900 (L) x 2000 (D) x 2200 (H) mm (with optional Stool installed on the right) |
|
Weight without the Scan ceph |
210 kg |
|
Weight with Scan Ceph |
240 kg |
Minimum Computer System Requirements
X-Ray Dose Emission Information
Imaging Performance Information
CS 9600 Environmental Requirements