Installing the Unit
Before you install the unit, carefully unpack and check that you have all the:
- Necessary tools
- Cables
- Components
 |
IMPORTANT: The unit is provided with a wall framework that by default requires an installation that orientates to the left. |
Preparing for Installation
For all installations, you must install the control timer unit at a place where:
- You can see:
- The selected parameters on the control timer unit.
- The patient from where you operate the control times unit.
- There is a minimum distance of 2.5 m between you and the patient.
 |
CAUTION:
- Make sure that you have the required installation cables.
- Check that the voltage displayed on your generator label corresponds to your power supply voltage.
- Check that jumper positions on the wall framework main powerboard matches the used voltage.
|
Related Topics
Unit Installation
Installing the Main Powerboard
Installing the CJ718 Powerboard
Connecting the Wiring
Post-Installation Control
Completing the Installation