Using the CS 8100 and CS 8100 3D Units or CS 8200 and CS 8200 3D Units
To prepare a unit:
Turn on the PC.
On the unit column, press the ON button and wait for the connection between the unit and the PC to be established.
Start the imaging software for the patient. Select the Acquisition icon to acquire a panoramic image (F3) and follow these steps:
Click the Program button to access the Program pane. Click on the section of the jaw to select the anatomical zone.
Note: For a TMJ x4 image, click the x2 button twice.
Click the Patient button to access the Patient pane. Select the patient type, size, and dental arch morphology.
Click the Parameter button to access the Parameter pane. Set the following parameters:
70 kV – 6.3 mA
80 kV – 10 mA
85 kV – 10 mA
If the default parameter settings
are not adapted to your patient type, click the Parameter
button, and in the Parameter
pane, select the appropriate parameters. To save the new settings
as the default, click and select Memorize
anatomy setting.
Position the appropriate chin rest on the chin rest support and cover the bite block with a hygienic barrier. If needed, use the edentulous bite block.