Using the CS 8100 and CS 8100 3D Units or CS 8200 and CS 8200 3D Units
Carry out the following daily maintenance procedures:
Bite block—Sterilize with cold sterilization, or autoclave up to 134°C for 18 minutes, before the next patient is x-rayed.
Edentulous bite block—Sterilize with cold sterilization, or autoclave up to 134°C for 18 minutes, before the next patient is x-rayed.
Temple supports and chin rest—Clean with medical-grade 76% alcohol disinfectant before the next patient is x-rayed.
Outer covers of the unit—Wipe the unit with a soft, dry cloth at the end of each day's operation.
Warning: Do not use detergents or solvents to clean the outer covers of the unit.
An authorized service technician should perform a general inspection of the unit annually.
If you are using the CS 8100 3D unit, see the CS 8100 3D Safety, Regulatory and Technical Specifications (SM844) for more information. If you are using the CS 8200 3D unit, see the CS 8200 3D Safety & Regulatory Guide (SMA67) for more information.