Error Messages

The following table lists the error messages, their possible causes, and the actions to take to fix the problems.

Table 7–3   Error Messages

Error Message

Reference ERR_

Description

Possible Cause

Actions

Comments

E01

OPERATOR

Early release of X-ray exposure switch

The user released the X-ray exposure switch before the end of the exposure.

Quickly press on the selection knob to stop the alarm.

Press and hold the exposure switch until the end of the exposure. Wait until the bip ends before you release the switch.

If the X-ray exposure switch was released before the end of the exposure cycle, you may save or develop the radiography. The remaining exposure time is displayed on the control timer unit.

The X-ray exposure switch or its wiring is defective.

Check the wiring, the X-ray exposure switch, and the electrical contact points.

Replace the defective parts.

The error is displayed when the X-ray exposure switch is enabled,. It happens often and unexpectedly.

E02

XRAY_SWITCH

X-ray control enabled when switching ON the unit

The exposure switch or its wiring is defective.

Check the wiring, the exposure switch and the electrical contact points. Replace the defective parts.

It appears on a CS 2200 unit if you use a CS 2100 control timer unit as an interface to access the technician menus.

E03

X_RAY_SWITCH_SECURITY

X-ray control safety malfunction

CJ718 powerboard failure

Switch OFF and ON the unit. If the error persists, change the CJ718 powerboard.

This error is displayed when you switch ON the unit.

E04

EXPO_TIMEOUT_SECURITY

Exposure time safety malfunction

CJ718 powerboard failure

Switch OFF and ON the unit. If the error persists, change the CJ718 powerboard.

This error is displayed when you switch ON the unit.

E10

E11

E12

KV_REF_SYS

kV_REF_LOW

kV_REF_HIGH

kV reference error

Too low kV reference

Too high kV reference

CJ718 powerboard failure.

Incorrect calibration of the CJ718 powerboard.

Launch the auto-calibration procedure.

If the problem persists, change the CJ718 powerboard.

 

E13

KV_FEEDBACK_SYS

kV return error (no-load value different from 0)

kV return connection error

Defective connections/wirings on kV return (i.e. disconnected J6 connector)

Switch OFF and ON the unit. If the error persists, do the following:

- Check the wiring and electrical contact points.

- Replace any defective parts.

 

Generator failure (kV return)

- Disconnect the generator and use an ohmmeter to check that the value between the yellow and blue generator wires are between 11kohms and 15kohms

- Change the generator.

CJ718 powerboard failure

- Change the powerboard.

E14

E15

kV_FEEDBACK_LOW

kV_FEEDBACK_HIGH

Low kV return

High kV return

Long period of generator inactivity

Switch OFF and ON the unit. If the error persists, carry out tube seasoning procedures. See Performing the Tube Seasoning.

 

Incorrect CJ718 generator calibration

Do a PR 4 parameter autocalibration. See Resetting and Autocalibration.

Defective wiring

Check the wiring and electrical contact points.

Replace any defective parts.

CJ718 powerboard failure

Change the CJ718 powerboard.

E16

FMIN_SYS

No signal frequency inverter error

CJ718 powerboard failure

-Switch OFF and ON the unit. If the error persists, change the CJ718 powerboard

 

E17

E18

FMIN_LOW

FMIN_HIGH

Too low inverter frequency

Too high inverter frequency

CJ718 powerboard failure

Switch OFF and ON the unit. If the error persists, change the CJ718 powerboard.

 

E20

HEATING_VOLTAGE_SYS

Error in voltage measure of heating filament (no-load value different from 0)

Disconnected or poorly connected generator when running the auto-calibration procedure.

- Switch OFF and ON the unit. If the error persists, do the following:

- Check the wiring and electrical contact points.

- Change the defective wire.

 

E21

HEATING_VOLTAGE_LOW

Too low voltage for filament heating

Incorrect CJ718 powerboard calibration

Do a PR 4 parameter autocalibration. See Resetting and Autocalibration.

E22

HEATING_VOLTAGE_HIGH

Too high voltage for filament heating

CJ718 powerboard failure

Change the CJ718 powerboard.

E23

HEATING_CURRENT_SYS

Error in current for filament heating (no-load value different from 0)

E24

HEATING_CURRENT_LOW

Too low current for filament heating

Generator failure (X-ray tube heating filament circuit)

- Switch OFF and ON the unit. If the error persists, do the following:

- Do an Ohmmeter test (<1ohm) between the purple and blue generator wires.

-Change the generator

Commonly results from a broken X-ray tube filament

Defective wiring (X-ray tube heating filament circuit)

- Do an ohmmeter test between the blue and purple wires on CJ718 slot J2

- Change the defective components

 

CJ718 powerboard failure

Change the CJ718 powerboard

E30

POWER_ALARM

Problem with line voltage setting or filtering capacitor

Incorrect configuration of the CJ718 powerboard voltage

Check the CJ718 powerboard jumper configuration for adaptation to the adequate voltage (the components that are referenced as 1a, 1b, and 2 on the electrical schema).

If not adequate with main voltage, change the CJ718 powerboard.

 

The unit power supply is out of the voltage range for proper use.

Check the electrical installation with emphasis on power supply voltage and connections.

- 90 V -144 V for 110 V

- 207 V and 264 V for 230 V

CJ718 powerboard failure

Change the CJ718 powerboard

 

E40

WATCHDOG

Watchdog release problem

- Interferences

- Random display of error message

- Check for incorrect grounding installation

- Check low cables cross-section

- Verify if cables from other closeby sources are not causing interferences

- Other appliances installed on the same power line

- If an error is detected bring the installation back to conformity

 

E41

PROGRAM_INTEGRITY

Firmware integrity default

E42

PROGRAM_FAULT

Firmware functions default

- CJ718 powerboard failure

- Systematic display of error message.

- Change the CJ718 powerboard

E43

SHOT_PARAMETERS

Last exposure parameter (exposure time)

E44

TECHNICAL_DATA

Technical parameters (settings)

E45

GENERATOR_COOLING_DATA

Tube head parameters (cooling)

E46

USER_DATA

User settings (digital, long cone)

E47

SW2_CONFIG

Incorrect configuration of CJ718 powerboard SW2 dipswitch 2

Incorrect configuration of CJ718 powerboard SW2 dipswitch 2

Modify SW2 dipswitch 2 configuration

 

E48

ERR_QUARTZ

Missing or unstable processor frequency

CJ718 powerboard failure

- Systematic display of error message.

- Switch OFF and ON the unit. If the error persists, change the CJ718 powerboard

 

E50

I2C

I2C bus

Defective wirings/connections

- Change the connections and the aspect of the electrical links.

- Change the defective parts.

When you hear a beep from the buzzer (beeps for 1 sec) and if no error code is displayed on the control timer unit, the problem surely results from either of the 3 problem causes listed in the previous Possible Causes column.

E52

I2C_EXT

External I2C bus

Interferences

- Random display of error message

- Check for incorrect grounding installation

- Check low cables cross-section

- Verify if cables from other closeby sources are not causing interferences

- Other appliances installed on the same power line

- If an error is detected bring the installation back to conformity.

E54

IHM

Error in accessing the control panel

CJ718 powerboard or control panel failure

- Systematic display of error message.

- Change the CJ718 powerboard.

- Change the control panel.

E51

I2C_INT

Internal I2C bus

E53

EPROM

Error in serial EPROM access

Related Topics

Maintenance

Preventive Maintenance

Corrective Maintenance

LEDs and Test Points