Using the CS 9600 Unit > Acquiring 2D Sinus Images
To prepare and position an adult or pediatric patient:
Ask the patient to remove all metal objects.
Ask the patient to wear a lead apron. Be sure the apron is flat across the patient’s shoulders.
On the touchscreen control panel,
press and hold the height adjustment buttons
to adjust the unit to the height of the patient. Ask the patient to
enter the unit.
Note: If the patient is tall, ask her to sit on a stool and then adjust the height.
Ask the patient to:
Stand up straight.
Grip the lower handle and slide hands together, palms up.
Rest the chin on the chin rest support and bite into the bite block.
Position the feet slightly forward.
Relax and lower the shoulders for full motion of the unit rotative arm.
Note: Correct posture reduces the shadow of the spinal column transferred to the image.
Position the patient:
For the Scanning area of interest (maxillary sinus exam):
Immobilize the patient’s head with the temple supports. Tighten the supports using the adjustors.
Select the area of interest.
For the Waters area of interest:
Tilt the head back.
Immobilize the patient’s head with the temple supports. Tighten the supports using the adjustors.
Select the area of interest.
For a Lateral R, Lateral L, AP, or PA area of interest:
Immobilize the patient’s head with the temple supports. Tighten the supports using the adjustors.
Select the area of interest.
When you see , use the
buttons on the control
panel to place the sensor as close to the patient as possible.
Ask the patient to:
Close the eyes.
Remain still.
Swallow and place the tongue against the palate.
Breathe through the nose.