Cleaning and Disinfecting the CS 9600
Precautions before use
Perform the following Cleaning and Disinfection activities on your product and accessories before use.
Cleaning the CS 9600 Unit and Accessory with No Patient Contact
To clean the unit and accessory with no patient contact such as the 3D bite block support, follow these steps:
- Switch off the unit.
- Remove all visible soil, if any, with a disposable cloth or paper wipe.
 |
Note: Do not disassemble unit parts. |
- Dampen (not soak) a lint-free cloth with soap and running water.
- Thoroughly clean manually all accessible parts of the unit, including the temporal head clamps and, if applicable, the 3D bite block support with the dampened lint-free cloth.
- Dry with a hygienic disposable cloth.
- Dampen (not soak) a lint-free cloth with a low-level disinfectant that is U.S. Environmental Protection Agency (EPA)-registered or a low-level disinfectant that is recognized by your Local Authority (for example, quaternary ammonium compounds and some phenolics). An EPA-registered hospital disinfectant or any other low-level disinfectant must have clear label claims for intended use.
- Wipe thoroughly on all accessible parts of the unit with the dampened lint-free cloth. You must follow the disinfectant’s manufacturer instructions for use, especially with respect to contact time.
- Allow the unit to dry in the open air for a minimum of 5 minutes.
- Visually inspect the unit for signs of deterioration. If any damage is noted, do not use the unit and contact a service technician.
 |
CAUTION: Avoid applying any cleaning liquid to the inside parts of the unit. |