Configuring the Monitor

The anesthesia records module interfaces with the following monitors:

The monitor must be configured to communicate with the anesthesia records module. The menu configurations for these monitors differ slightly; however each requires the same settings: Print Device, Serial Format, and Baud Rate. In addition, each vital sign being monitored must be set to ON using the PARAMS menu.

Click your monitor for configuration settings:

POET PLUS 8100 Vital Signs MonitorPOET PLUS 8100 Vital Signs Monitor

Ensure the following settings are configured on the CONFIG menu:

Print Device - Serial

Serial Format - CUSP

Baud Rate - 38400

Ensure the following settings are configured on the PARAMS menu:

SPO - ON

RESP - ON

TEMP1 - ON

TEMP2 - ON

Unit of Measure - Degrees F

 

NGENUITY 8100E(P) Patient MonitorNGENUITY 8100E(P) Patient Monitor

Ensure the following settings are configured on the PRINT menu:

Print Device - Serial

Serial Format - CUSP

Baud Rate - 38400

Ensure the following settings are configured on the PARAMS menu:

SPO - ON

RESP - ON

TEMP1 - ON

TEMP2 - ON

Unit of Measure - Degrees F

 

506N3 Vital Signs Monitor and 507ELC2 Scholar III Vital Signs Monitor506N3 Vital Signs Monitor and 507ELC2 Scholar III Vital Signs Monitor

Ensure the following settings are configured on the Printer Setup menu:

Print To - SERIAL

Serial - CUSP

Baud Rate - 19200

Ensure the following settings are configured on the Configuration menu:

NIBP - ON

SPO2 - ON

TEMP - ON

PPM3PPM3

The following settings must be configured, and then the monitor must be powered off and on to enable the settings:

From the SETUP > ADMINISTRATION > COMMUNICATIONS menu:

Connect Type - Serial

From the SETUP > ADMINISTRATION > CONNECTIVITY menu:

EMR Interface Type - 1

SurveyorSurveyor

The following settings must be configured, and then the monitor must be powered off and on to enable the settings:

From the SETUP > ADMINISTRATION > COMMUNICATIONS menu:

Connect Type - Serial

From the SETUP > ADMINISTRATION > CONNECTIVITY menu:

EMR Interface Type - 1

MindrayMindray

Connecting the Mindray Monitor

Contact a Carestream Dental Support Representative.

 