| Parameters | |
|
IEC standard 60601-2-65 |
Conform |
|
Type of protection against electric shocks |
Class 1 |
|
Degree of protection against electric shock |
Type B |
|
Rated value of inherent filtration |
1.0 mm (0.039 in.) eq. Al |
|
Rated value of total filtration |
At least 1.5 mm (0.059 in.) eq. Al |
|
Beam-limiting cone, focal spot/skin distance |
20 cm (8 in.) |
|
Maximum accumulated heat |
32.500 J |
|
Maximum continuous thermal dissipation |
7 W at 70 kV |
|
Leaking radiation at maximum rate during one hour’s use |
< 0.25 mGy |
|
Maximum field of symmetrical radiation |
6 cm (2.375 in.) diameter |
|
Position and tolerance of the focal point on the reference axis |
0 mm +/- 0.5 mm (0.020 in.) |
Figure 6 X-Ray Source Assembly

Figure 7 Heating and Cooling Curves of CS 2100 and CS 2200 X-Ray Source Assembly