CS 2100 | CS 2200 | |||
Type of electrical power supply |
220-240 V ~ (±10 %), 50/60 Hz |
100-130 V ~ (± 10 %), 50/60 Hz |
220-240 V ~ (±10 %), 50/60 Hz |
100-130 V ~ (± 10 %), 50/60 Hz |
Apparent resistance of the power supply circuit |
0.5 Ω (max) |
0.2 Ω (max) |
0.5 Ω (max) |
0.2 Ω (max) |
Permanent absorbed current |
220 mA |
|||
Current absorbed during the X-ray emission |
5 A |
12 A |
5 A |
12 A |
Nominal high voltage |
60 kV |
70 kV |
||
Maximum corresponding tube current |
7 mA |
7 mA |
||
Nominal tube current |
7 mA |
7 mA |
||
Maximum corresponding high voltage |
60 kV |
70 kV |
||
Tube current/voltage combination for maximum output power |
60 kV, 7 mA |
70 kV, 7 mA |
||
Nominal power for an exposure time of 0.1 s |
420 W at 60 kV, 7 mA |
490 W at 70 kV, 7 mA |
||
Minimum value of the current/time in the range of conformity |
0.07 mAs at 7 mA |
0.07 mAs at 7 mA |
||
Maximum Use Rate |
at 60 kV, 7 mA, 0.1 s and at the maximum X-ray source assembly: approximately one exposure every 6 s |
at 70 kV, 7 mA, 0.1 s and at the maximum X-ray source assembly: one exposure every 7 s |
||
Selection of the Load Parameters |
Fixed parameters: 60 kV, 7 mA
|
70 kV, 7 mA and 60 kV, 7 mA |
Measurement Conditions | |
kV |
Indirect measurement with the peak kilovolt meter |
mAs |
Direct in the circuit with a mAs-meter |
Exposure time |
Indirect measurement on the kV signal at 75 % of the peak value |
Accuracy of the Load Parameters | |
High voltage |
kV ±10% |
Current in the tube |
mA ±20% |
Exposure time seconds |
Seconds ±(10% + 1ms) or ±(5% + 50 ms) |