Pediatric Use: Summary of Safety Instructions
By reducing the Field of View for 3D X-ray imaging on children or adolescents, you are reducing the exposed area and this reduces the dose received by the patient.
For a recommendation on the Field of View (FoV) to be selected for children/adolescents, see the table below:
Field of View Selections
Standard Field of View |
Recommended Field of View for Children/Adolescents |
5x5 |
4x4 |
6x6 |
5x5 |
10x5 |
8x5 |
10x10 |
8x8 |
12x5 |
10x5 |
12x10 |
10x10 |
16x17 |
16x12 |