Occasionally, malfunctions can occur during use in the event of an incorrect action. An error message is displayed on a pop-up window on the Acquisition interface System Status screen.
The following table lists the information messages, their description, and the action to take:
![]() |
IMPORTANT:
|
Table 2 Error Code Message
|
Error Code |
Error Message |
Description |
Action |
|
Err_S_GEN_36865 |
Handswitch was released before the end of exposure. |
The user has released the exposure button prematurely. |
Relaunch the acquisition and hold the exposure button until the end of the acquisition cycle. |
Table 3 Quick Troubleshooting
|
Message |
Description |
Action |
|
X-ray tube cooling |
Cooling in progress. |
Wait until the Generator Cooling Indicator on the Acquisition interface reaches zero. |
|
Thermal security |
Cooling in progress. |
Wait until the Generator Cooling Indicator on the Acquisition interface reaches zero. |
|
Release handswitch |
The acquisition has ended. |
Release the exposure button of the X-ray remote control. |
|
Start the acquisition |
The acquisition has been launched. |
Keep pressing and holding the exposure button. |
|
Upgrade of the firmware |
The system is being upgraded. |
Wait for completion of the upgrade. |