Site Preparation before Installation

important_icon IMPORTANT: Prior to placing the order and before installation, you must carefully check the following requirements.

Standard Compliance

The compliance with all national and local regulations and with the manufacturer’s recommendations is mandatory for high and low voltage wiring.

Environmental Requirements

Check the following ambient operating condition requirements of the X-ray room before installing the unit:

Temperature: 5~35 °C

Relative humidity: 30~85%

Atmospheric pressure: 700~1060 hpa

Mechanical Requirements

warning_icon WARNING: The CS 2200 Intraoral X-ray unit must be installed so that it is impossible to rotate the scissor arm through 360°. Otherwise, the power supply cable can fail and cause electrical damage.
important_icon IMPORTANT: To attach the wall framework to the wall, you must use an appropriate fixing system depending on the type of wall. It is the responsibility of the installer to choose the appropriate fixing system which must be able to withstand an extraction force of 147 kg (324 lbs) at each point of attachment.